EasyManua.ls Logo

Horstmann ComPass B - Page 17

Horstmann ComPass B
29 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use
ComPass B 4023841-001
ComPass B menu settings for voltage calibrationStep 4:
Select menu:
Settings →↓ Expert Mode →↓ Grid parameter
- Set frequency to 50 Hz
- Set VNom to 11.0 kV (Voltage used in this example !)
Select menu:
Settings →↓ Expert Mode →↓ V-Calibration
Set VCAL_LL to 11000V.
Note: VCAL_LE entry is changed automatically.
G
Alternatively, set VCAL_LE parameter whereupon
VCAL_LL is changed automatically.
Select entry „Start calibration“ and push rocker switch to the
right.
If the following security prompt appears, V-AutoCal?
=> Select “Yes”.
Note: Presence of calibration voltage of 11 kV/6.35 kV is requi-
red!
Indication on the display during the calibration process.
Indication on the display after calibration has been completed.
Correction values are stored.
Push rocker switch to the left!
Step 5:
Indication:
- Voltage is calibrated.
- Date of calibration.
- Calibration valid for medium-voltage range.
Verification if calibration has been successfully completed:
Applied calibration voltage shall be identical with the voltage va-
lue indicated on the display (to be found in the menu under mea-
surement results).
17
Dipl.-Ing. H. Horstmann GmbH