.6 Set the Calibrator for the first value listed in the Applied column of Table 8 and set the OPR/STBY switch to
OPR.
4
.4
.7 Edit the Calibrator controls until the AC Measurement Standard indicates the value listed in the Applied
column of Table 8.
4
.4
.8 The TI display must indicate within the corresponding values listed in the Limits column of Table 8.
4
.4
.9 Set the Calibrator OPR/STBY switch to STBY.
4
.4
.10 Repeat steps 4
.4
.4 through 4
.4
.9 for each remaining range and applied value listed in Table 8.
NOTE
Grounding/Shielding variations may be required to obtain a stable and accurate
TI indication
. The TI LFILTER may be used in Table 8 for frequencies up
through 50 kHz . Disengage TI LFILTER for frequencies above 50 kHz
. To
engage TI LFILTER press the TI keys in the following sequence:
1
Eller
Table 8.
Range
(AC)
Applied (VAC)
Limits (VAC)
10 mV
10m@20Hz
9
.99215 to 10
.00785 m
l0m@1kHz
9
.99532 to 10
.00468 m
10m@20kHz
9
.99440 to 10
.00560 m
10m@50 kHz
9
.98668 to 10
.01332 m
10m@ 100kHz
9
.94630 to 10
.05370 m
l00 mV
100m@20Hz
99
.9835 to 100 .0165 m
100m@1kHz
99
.9885 to 100
.0115 m
100m@20kHz
99
.9820 to 100 .0180 m
100m@50 kHz
99
.9652 to 100
.0348 m
l00 m @ 100 kHz 99
.9139 to 100
.0861 m
1V
1 @20 Hz
0
.999864 to 1
.000136
1 @ 1 kHz *)
0
.999906 to 1 .000094
1 @ 20 kHz
0
.999838 to 1
.000162
1 @50kHz
0
.999676 to 1 .000324
1 @ 100 kHz
0
.999177 to 1 .000823