Table 8
. (Cont)
Range
(AC)
Applied (VAC) Limits (VAC)
10V
10 @20 Hz
9
.99863 to 10
.00137
10 @ 1 kHz
9
.99905 to 10
.00095
10@20 kHz
9
.99837 to 10
.00163
10@50 kHz
9
.99676 to 10
.00324
10 @ 100 kHz
9
.99176 to 10
.00824
100 V
100 @20 Hz
99
.9749 to 100
.0251
100 @ 1 kHz
99
.9778 to 100
.0222
100@20 kHz
99
.9778 to 100
.0222
100 @ 50 kHz
99
.9623 to 100
.0377
100 @ 100 kHz
99
.8776 to 100
.1224
1000 V
200 @ 40 Hz
199
.900 to 200
.100
300 @ 10 kHz 299
.800 to 300.200
500 @ 100 kHz 498
.459 to 501
.541
1000 V
(Opt HO1 only)
1k@30 kHz
*2
998
.773 to 1001
.227
*
I
Record the TI indication.
*2
If required, manually range the TI and AC Measurement Standard.
4
.4
.11 Disconnect the test setup.
4
.4
.12 Set the TI to the first range listed in the Range column of Table 9.
4
.4
.13 Connect the Calibrator WIDEBAND output, using the cable and 50 S2 termination supplied with the Calibrator,
to the AC Measurement Standard INPUT 2.
4
.4
.14 Set the AC Measurement Standard to measure the first value listed in the Applied column of Table 9.
4
.4
.15 Set the Calibrator for the first value listed in the Applied column of Table 9 and set the OPR/STBY switch to
OPR.
4
.4
.16 Edit the Calibrator controls until the AC Measurement Standard indicates the value listed in the Applied
column of Table 9.
4
.4
.17 Set the Calibrator OPR/STBY switch to STBY.
4
.4
.18 Disconnect the AC Measurement Standard from the Calibrator WIDEBAND output
.
27