HP 3458A - AC Voltage Calibration (Synchronous Sub-sampled Mode)

HP 3458A
43 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
T
.O
. 33K8-4-551-1
Table 8
. (Cont)
Range
(AC)
Applied (VAC) Limits (VAC)
10V
10 @20 Hz
9
.99863 to 10
.00137
10 @ 1 kHz
9
.99905 to 10
.00095
10@20 kHz
9
.99837 to 10
.00163
10@50 kHz
9
.99676 to 10
.00324
10 @ 100 kHz
9
.99176 to 10
.00824
100 V
100 @20 Hz
99
.9749 to 100
.0251
100 @ 1 kHz
99
.9778 to 100
.0222
100@20 kHz
99
.9778 to 100
.0222
100 @ 50 kHz
99
.9623 to 100
.0377
100 @ 100 kHz
99
.8776 to 100
.1224
1000 V
200 @ 40 Hz
199
.900 to 200
.100
300 @ 10 kHz 299
.800 to 300.200
500 @ 100 kHz 498
.459 to 501
.541
1000 V
(Opt HO1 only)
1k@30 kHz
*2
998
.773 to 1001
.227
*
I
Record the TI indication.
*2
If required, manually range the TI and AC Measurement Standard.
4
.4
.11 Disconnect the test setup.
4
.4
.12 Set the TI to the first range listed in the Range column of Table 9.
4
.4
.13 Connect the Calibrator WIDEBAND output, using the cable and 50 S2 termination supplied with the Calibrator,
to the AC Measurement Standard INPUT 2.
4
.4
.14 Set the AC Measurement Standard to measure the first value listed in the Applied column of Table 9.
4
.4
.15 Set the Calibrator for the first value listed in the Applied column of Table 9 and set the OPR/STBY switch to
OPR.
4
.4
.16 Edit the Calibrator controls until the AC Measurement Standard indicates the value listed in the Applied
column of Table 9.
4
.4
.17 Set the Calibrator OPR/STBY switch to STBY.
4
.4
.18 Disconnect the AC Measurement Standard from the Calibrator WIDEBAND output
.