T
.O
. 33K8-4-551-1
Table 1
. (Cont)
Test Instrument (TI)
Performance
Test
Characteristics
Specifications
Method
DC Voltage
(Cont
.)
Range
: 0 to 1000 VDC
Compared to a DC
Reference Standard
Accuracy
:
and DC Reference
Relative
: ±(ppm of rdg + ppm of mg)
Divider or DC Voltage
Absolute
: ±[Sqrt(TI ppm of rdg
A
2 +
Divider
STD 1 ppm of rdg^2 + STD2 ppm of rdg^2)
+ TI ppm of mg]
Opt 002
(Cont
.)
Range
Relative
Absolute
100 V
6 + 0
.3
(6 + 1
.5 + 0
.2) + 0
.3
1000 V
6 + 0
.1
(6 + 1
.5 + 0
.5) + 0
.1
Inputs >100 VDC, add
12 ppm X (Vinll000)
2
Resistance
Range
Range
: 0 to 1 G52
Compared to a
Resistance Standard
Accuracy:
Relative
: ±(ppm of rdg + ppm of mg),
(Add 250 mf
g
for Two-Wire Ohms)
Absolute: ±[Sgrt(TI ppm of rdg^2 +
STDI ppm of rdg^2) + Ti ppm of mg]
(Add 250 ml) for Two-Wire Ohms)
Relative
Absolute
10 S2
15+5
(15+2
.15)+5
10052
12+5
(12+2
.15)+5
1 kl2
10+0
.5
(10+2
.15)+0
.5
10ki2
10+0
.5
(10+2.1)+0
.5
100kI2
10+0
.5
(10+2.4)+0
.5
1 MSZ
15+2
(15+6)+2
10 MCI
50+10
(50 + 12) + 10
100 MS2
500 + 10
(500 + 50) + 10
1 G1)
5000 + 10
(5000 + 50) + 10