T
.O
. 33K8-4-551-1
Table 1
. (Cont
.)
Test Instrument (TI)
Performance
Characteristics
Specifications
Test
Method
DC Current
Range
: 0 to 1 ADC
Compared to a known
DC Current
Accuracy:
Relative
: ±(ppm of rdg + ppm of mg)
Absolute
: ±[Sgrt(TI ppm of rdg^2 +
STD1 ppm of rdg^2 + STD2 ppm of rdg^2)
+ TI ppm of mg]
Range
Relative
100 nA
*
I
30 + 400
1 µA
*
I
20 + 40
10µA *'
20+10
100 µA
20 + 8
1 and 10 mA
20 + 5
100mA
35+5
lA
110+ 10
Absolute
N/A
N/A
N/A
(20 + 5
.277 + 5) + 8
(20 + 5
.277 + 5) + 5
(35+5
.277+5)+5
(110+5
.277+6)+10
AC Voltage
Range
: 0 to 1000 VAC
(Synchronous Sub-sampled Mode)
AC BAND <_2 MHz
Compared to an
AC Measurement
Standard or Calibrator
*2
Accuracy:
Relative
: ±(% of rdg + % of mg)
Absolute:
±[Sgrt(TI % of rdg^2 + STD1 % of rdg^2 +
STD2 % of rdg^2) + TI % of mg +
STD1 µV + STD2 [IV]
Range
Frequency
Relative
Absolute
10 mV
1 to 20 Hz
0
.03 + 0
.03
(0
.03+0+0
.029)+0
.03+0+ 1
.3
20 to 40 Hz
0
.03 + 0
.03
(0
.03+0+0
.019)+0
.03+0+ 1
.3
40 Hz to 1 kHz
0
.02 + 0
.011
(0
.02 + 0 + 0
.011) + 0
.011 + 0 + 1
.3
1 to 20 kHz
0
.03 + 0
.011
(0
.03 +0+0
.011)+0
.011 + 0 + 1
.3
20to50kHz
0
.1+0
.011
(0
.1+0+0
.021)+0
.011+0+2
50 to 100 kHz
0
.5 + 0
.011
(0
.5+0+0
.031)+0
.011+0+2
.5
100 to 300 kHz
4 + 0
.02
(4+0+0
.081)+0
.02+0+4
See footnotes at end of Table
.