Section V
5.5. PEBFOBMANCE  CHECKS.
5-6. 
The 
following 
performance 
checks  are in{abinet
procedures 
that 
can be used to verify that the Model 
3575A
meets the specifications listed in Table 
1-1. 
These 
proce-
dures 
can 
be used for 
incoming 
quality 
control  inspection,
to 
check specifications  after a 
repair or 
for 
routine
maintenance.
5-7. Test Card.
5-8. A Performance  Check Test Card 
is 
provided 
at the 
end
of this section 
for 
your 
convenience 
in 
recording  the
performance of the Model  35754 during 
performance
checks, This card can be 
removed from the manual and
used 
as 
a 
permanent record of 
the 
incoming  inspection 
or
of a routine 
performance 
check. The test 
card may be
reproduced 
without written 
permission from 
Hewlett-
Packard.
$9. 
Panel  Meter Accuracy Check.
5-10. 
The 
purpose 
of this check is to 
yerify 
that 
the 
panel
meter accuracy is within the 
i 
3 count 
(1 
0.3 
dB,
t 
0.3 
deg.) tolerance  listed in Table 1-1.Ifthe 
panel 
meter
fails to 
meet the required specifications, 
perform 
the 
Panel
Meter 
Accuracy  Adjustments 
(Paragraph 
5-29) before 
pro-
ceeding with the other 
perforrnance 
checks 
in 
this 
section.
RTCOMMENDED TEST EQUtrMENT:
DC Digital 
Voltmeter 
(-hp- 
Model  3450A)
Frequency  Synthesizer 
(hp- 
Model 
3320B)
Variable Attenuator 
(-hp- 
Model  355D)
(2) 
50-Ohm Feed-Thru Terminations 
(hp- 
Model
l 1048C)
50-Ohm  Pow 
Splitter 
(See 
Figure  5-l)
a. Connect  test equlpment
Conneot  DC Digital Voltmeter
Model 
35754
as shown in Figure  5-2.
to ANALOC OUTPUT 1.
b. 
Set the 
3575A controls 
as 
follows:
DISPLAY*  . .. AMPLITUDE
AMPLITUDE 
FUNCTION.  .. . .. . . B/A
Voltage Range  - 
0.2 mV 
- 
2 V
(both 
channels)
FREQUENCY  RANGE ....l 
[lz 
- 
1kIlz
PHASEREFERENCE  ........,....  A
* 
Options 
U01 
003: 
Set 
AMPLITUDE
B/PHASE 
swilch 
to 
PHASE.
c. 
Set the 
Frequency 
Synthesizer amplitude  to
+ 
25.05 dBm 
(2 
V 
rms 
at outputs of Power 
Splitter) and
frequency 
to I kl1z.
d. Set the Variable 
Attenuator  to 0 
dB.
e. 
Allow 
30 seconds 
for 
the 
readings to 
stabilize.
f.  Observe  the DC Digital Voltmeter reading and the
B/A 
panel 
meter reading. 
The 
panel 
meter reading  should
be within 
t 
3 caunts 
of 
the 
DVM reading  e.g., 
the analog
output voltage 
is 10 
mV 
per 
dB. 
An 
output of 
+ 
I mV dc
should, therefore, 
produce 
a 
panel 
meter reading 
of
+ 
00.1 dB 
t 
0.3 dB 
(3 
counts).
9., 
Set 
the Variable 
Attenuator 
to 
80 
dB and allow  30
seconds 
for 
the 
readings 
to stabilize.
h. 
Observe the 
DVM reading 
and the B/A 
panel 
meter
reading. 
The 
panel 
metff 
reading 
should 
be within 
t 
3
counts of 
the 
DVM reading  e.g., if the DVM reading is
- 
0.797 
Ydc, the 
panel 
meter 
reading 
should  be 
- 
79.7 
dB
r 0.3 dB.
PEBFOBMANCE  CHECKS
Parts 
Lisr.
Designator
-hp- 
Part No.
Jl 
,J3
R1 
.R3
Eoard
Connector: BNC 
lemale
R I 
f 
xd flm 16.9 ohms 
'l
1.5"X1.5"Copper-Clad
%112W
(double'sided)
1250-0083
0698-5005
Figure 5-1. Power 
Splitter.