EasyManuals Logo

HP 53132A Service Manual

HP 53132A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #122 background imageLoading...
Page #122 background image
Chapter 2 Service
About the HP 53181A Calibration Menu
2-26 Assembly-Level Service Guide
2
unsecure for calibration by entering the security code while the
Counter is secured (CODE: )
change the security code by entering a new code while the Counter is
unsecured (CODE: )
view the calibration count, which indicates the number of times that
the Counter has been calibrated (CAL COUNT?). The calibration
count increments up to a maximum of 32,767 after which it wraps
around to 1. A calibration count of 0 indicates that the Counter is
completely uncalibrated. (The calibration count is unaffected by
interpolator calibration.)
get Help (HELP:) with any of the following:
how to calibrate (CAL?),
how to secure (SECURE?), and
how to change the code (CODE?)
NOTE
A good precaution is to use an HP-IB program (see the programming
example titled “To Read and Store Calibration Data” in Chapter 3 of the
Programming Guide) to read and store the calibration factors prior to
initiating any calibration.

Table of Contents

Other manuals for HP 53132A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the HP 53132A and is the answer not in the manual?

HP 53132A Specifications

General IconGeneral
BrandHP
Model53132A
CategoryCash Counter
LanguageEnglish

Related product manuals