EasyManua.ls Logo

HP 53132A - Test 1: Time Interval (HP 53131 A;132 A Only); Equipment; Counter Setup

HP 53132A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Performance Tests
Test 1: Time Interval (HP 53131A/132A Only)
Assembly-Level Service Guide 1-17
1
Test 1: Time Interval (HP 53131A/132A Only)
This test verifies the Time Interval specifications between Channels 1 and 2.
Equipment
HP 8130A Pulse Generator (or equivalent)
HP 10503A BNC Cable
Counter Setup
1Cycle the POWER key to preset the Counter.
2Press Time & Period key until
TI 1 TO 2
is displayed.
3 Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
AUTO TRG: OFF
Trigger/Sensitivity,
LEVEL: .500V (Enter)
Trigger/Sensitivity,
COMMON 1: ON
50
/1M
, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
4 Set CHANNEL 2 input conditions to:
Trigger/Sensitivity,
AUTO TRG: OFF
Trigger/Sensitivity,
LEVEL: .500V (Enter)
Trigger/Sensitivity,
SLOPE: NEG
50
/1M
, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off