EasyManua.ls Logo

HP 53132A - Test 4: Channels 1 and 2 Frequency Accuracy (HP 53131 A;132 A Only); Equipment

HP 53132A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 1 Performance Tests
Test 4: Channels 1 and 2 Frequency Accuracy (HP 53131A/132A Only)
Assembly-Level Service Guide 1-27
1
Test 4: Channels 1 and 2 Frequency Accuracy
(HP 53131A/132A Only)
This set of tests verifies the frequency accuracy specifications of the
HP 53131A and HP 53132A Universal Counters.
Equipment
HP 8663A Synthesized Signal Generator
HP 1250-0780 Type N-to-BNC Connector
HP 10100C 50
Feedthrough
HP 10503A BNC Cables (2)
Counter Setup
1Cycle the POWER key to preset the Counter.
2 Using the Gate & ExtArm key, set the gate time to
TIME 1.000 s
.
3 Set CHANNEL 1 input conditions to:
Trigger/Sensitivity,
AUTO TRG: OFF
Trigger/Sensitivity,
LEVEL: 0.000 V
50
/1M
, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off
4 Set CHANNEL 2 input conditions to:
Trigger/Sensitivity,
AUTO TRG: OFF
Trigger/Sensitivity,
LEVEL: 0.000 V
50
/1M
, 50
DC/AC, DC
X10 Attenuate, Off
100kHz Filter, Off