Chapter 1 Performance Tests
Test 7: Peak Volts, Channels 1 and 2 (HP 53131A/132A Only)
Assembly-Level Service Guide 1-39
1
Volt Peak 2
1 Remove the signal from Channel 1 of the Counter and connect it
to Channel 2.
2Press Other Meas key until
VOLT PEAK 2
is displayed.
The Counter should display -1.00 ±0.12V for the negative peak of the sine
wave. This reading is displayed on the left side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 2a).
Also, the Counter should display +1.00 ±0.12V for the positive peak of the
sine wave. This reading is displayed on the right side of the display.
Record the actual reading in the Performance Test Record
(Test 7, Line 2b).
3 Disconnect the test setup.