Chapter 2  Service
The HP 53181A Calibration Procedures
2-32 Assembly-Level Service Guide
2
c. Press   key once, and then press   key three times.
The Counter should display 
CODE: 53000
.
d. Press   key once, and then press   key once.
The Counter should display 
CODE: 53100
.
e. Press   key once, and then press   key eight times.
The Counter should display 
CODE: 53180
.
f. Press   key once, and then press   key once.
The Counter should display 
CODE: 53181
.
g. Finally, press Enter key.
CAL UNSECURE
 is displayed.
NOTE
If “53181” is NOT the correct security code, the Counter will display 
BAD CODE
.
4Press Scale & Offset key until 
CAL: OFFS 1 ?
 is displayed.
5Press Enter key.
A scrolling message on the display will ask you to remove any signals 
connected to Channel 1. Follow the instruction.
6Press Enter key.
The Counter momentarily displays 
CALIBRATING
, and then it should 
display 
OFFS 1 PASS
.
If the fail message is displayed, refer to the troubleshooting section in this 
chapter.
This completes the Offset Calibration procedure. Proceed to the next 
procedure to calibrate the gain.