EasyManuals Logo

HP 53132A Service Manual

HP 53132A
254 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #227 background imageLoading...
Page #227 background image
Chapter 7 HP 53131A/132A Specifications
Measurement Specifications (Continued)
Assembly-Level Service Guide 7-9
7
Time Interval
Measurement is specified over the full signal ranges
7
of Channels 1 and 2.
Results Range: –1 ns to 10
5
s
LSD:
RMS Resolution:
t
res
Systematic Uncertainty: ±(Time Base Error × TI) ± Trigger Level Timing Error ± 1.5 ns Differential Channel Error (HP 53131A)
± 900 ps Differential Channel Error (HP 53132A)
Time Interval Delay
After a Time Interval Measurement has begun by satisfying the trigger conditions on Channel 1, the instrument will wait for the
user-entered delay time to elapse before the end-of-measurement trigger will be accepted on Channel 2. Please refer to Measurement
Arming for additional information.
Frequency Ratio:
Measurement is specified over the full signal range of each input.
Results Range: 10
–10
to 10
11
‘Auto’ Gate Time: 100 ms (or sufficient cycles on Channel 2 or 3 to make a valid measurement, whichever is longer)
LSD:
RMS Resolution:
To minimize relative phase measurement error, connect the higher frequency signal to Channel 1 when possible.
7
See Specifications for Pulse Width and Rise/Fall Time measurements for additional restrictions on signal timing characteristics.
Measurement Specifications (Continued)
53131A 53132A
500 ps 150 ps
--------------------------------------------------------------------------------------------
t
res
2
Start Trigger Error
2
Stop Trigger Error
2
++
53131A 53132A
750 ps 300 ps
--------------------------------------------------------------------------------------------
Ch1
Ch2
Ch1
Ch3
Ch2
Ch1
Ch3
Ch1
Ratio
1
2
:
1
Ch2 Freq Gate Time
×
-----------------------------------------------------------
Ratio
1
3
:
1
Ch3 Freq Gate Time
×
-----------------------------------------------------------
Ratio
2
1
:
Ch2 Freq
Ch1 Freq
()
2
Gate Time
×
------------------------------------------------------------------
Ratio
3
1
:
Ch3 Freq
Ch1 Freq
()
2
Gate Time
×
------------------------------------------------------------------
Ratio
1
2
:
2 1 Ch1 Freq Ch2 Trigger Error
×()
2
+×
Ch2 Freq Gate Time
×
-------------------------------------------------------------------------------------------------------------
Ratio
1
3
:
2 1 Ch1 Freq Ch3 Trigger Error
×()
2
+×
Ch3 Freq Gate Time
×
-------------------------------------------------------------------------------------------------------------
Ratio
2
1
:
2 Ch2 Freq 1 Ch1 Freq Ch2 Trigger Error
×()
2
+××
Ch1 Freq
()
2
Gate Time
×
--------------------------------------------------------------------------------------------------------------------------------------------
Ratio
3
1
:
2 Ch3 Freq
×
1 Ch1 Freq Ch3 Trigger Error
×()
2
+×
Ch1 Freq
()
2
Gate Time
×
--------------------------------------------------------------------------------------------------------------------------------------------

Table of Contents