Procedure 9. Al4 and A15 Assemblies
Procedure 9.
Al4 and Al5 Assemblies
Removal
1. Remove the spectrum analyzer cover as described in “Procedure 1. Spectrum Analyzer
Cover.”
2. Place the spectrum analyzer on its right side frame.
3. Remove the eight screws (1) holding the Al4 and Al5 assemblies to the bottom of the
spectrum analyzer. See Figure 4-18.
1
P
Figure 4-18. Al4 and A15 Assembly Removal
I
/
I
A)
1
s1112e
Caution
Washers are not captive. Loose washers in instrument may cause internal
damage.
Assembly Replacement 4-37