EasyManua.ls Logo

HP 8753D - Page 332

HP 8753D
678 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration Considerations
Measurement
P
arameters
Calibration
procedures
are
parameter-specic
, rather
than channel-specic
. When
a parameter
is
selected,
the
instrument
checks
the
available calibration
data, and
uses the
data found
for
that
parameter
.
F
or
example,
if a
transmission response
calibration is
performed
for
B/R,
and
an
S
11
1-port
calibration
for
A/R,
the
analyzer
retains
both
calibration sets
and
corrects
whichever
parameter
is
displayed. Once
a calibration
has been
performed
for
a
specic
parameter
or
input,
measurements of
that parameter
remain calibrated
in
either
channel,
as
long
as
stimulus values
are coupled.
In the
response
and
response
and
isolation
calibrations
,
the
parameter
must be
selected before
calibration: other
correction procedures
select
parameters
automatically
.
Changing channels
during a
calibration procedure
invalidates
the
part
of
the
procedure already
performed.
Device
Measurements
In
calibration
procedures
that
require
measurement
of
several
dierent
devices
,
for
example
a
short,
an
open,
and
a
load,
the
order
in
which
the
devices
are
measured
is
not
critical.
Any standard
can be
re-measured, until
the
N
N
N
N
N
NN
NN
NN
N
N
N
DONE
key
is pressed.
The change
in
trace
during
measurement
of a
standard is
normal.
Response
and response
and isolation
calibrations
require
measurement
of
only
one
standard
device
.
If
more
than
one
device
is
measured,
only
the
data
for
the
last
device
is
retained.
Omitting
Isolation
Calibration
Isolation
calibration
can
be
omitted for
most
measurements
,
except
where
high
dynamic
range
is
a
consideration.
Use
the following
guidelines
.
When
the
measurement
requires
a
dynamic
range
of:
90
dB:
Omit
isolation
calibration
for
most
measurements
.
90 to
100
dB:
Isolation
calibration
is
recommended
with
test
port
power
greater
than
0
dBm. F
or
this
isolation
calibration,
averaging
should
be
turned
on
with
an
averaging
factor
at least
four
times
the
measurement
averaging
factor
.
F
or
example
,
use
use
an
averaging
factor of
16 for
the
isolation
calibration,
and
then
reduce
the
averaging
factor
to
four
for
the
measurement
after
calibration.
100
dB:
Same
as
above
,
but
alternate
mode
should
be
used. See
page 5-53.
Saving Calibration Data
You should save the calibration data,
either in the internal non-volatile memory or on a disk.
If you do not save it, it will
be lost if you select another calibration procedure for the same
channel, or if you change stimulus values
. Instrument preset, power on, and instrument state
recall will also clear the calibration data.
Application and Operation Concepts 6-65