Each
standard must
be identied
as one
of
ve
\types":
open,
short,
load,
delay/thru,
or
arbitrary
impedance.
After
a standard
number
is
entered,
selection
of
the
standard
type
will
present
one
of ve
menus
for entering
the
electrical
characteristics
(model
coecients)
corresponding
to
that
standard type
, such
as
NN
NN
NN
NN
NN
NN
N
N
OPEN
.
These menus
are tailored
to the
current type
,
so
that
only
characteristics
applicable
to
the
standard
type
can
be
modied.
The
following
is
a
description
of
the
softkeys
located
within
the dene
standard menu:
N
N
N
N
N
N
N
N
N
N
N
NN
N
OPEN
denes
the
standard
type
as
an
open,
used
for calibrating
reection measurements
.
Opens are
assigned
a
terminal
impedance
of
innite
ohms
,
but
delay
and
loss
osets
may
still be
added.
Pressing
this
key
also
brings
up
a
menu
for
dening
the
open,
including
its
capacitance
.
As
a
reection
standard,
an
open
termination
oers
the
advantage
of
broadband
frequency
coverage
.
At
microwave
frequencies
,
however
,
an
open
rarely
has
perfect
reection
characteristics
because
fringing
capacitance
eects
cause
phase
shift
that
varies
with
frequency
.
This
can
be
observed
in
measuring
an
open
termination
after
calibration,
when
an
arc
in
the
lower
right
circumference
of
the
Smith
chart
indicates
capacitive
reactance
.
These
eects
are
impossible
to
eliminate
,
but
the
calibration
kit
models
include
the
open
termination capacitance
at
all
frequencies
for
compatible
calibration
kits
.
The
capacitance
model is
a cubic
polynomial,
as
a
function
of
frequency
,
where
the
polynomial
coecients
are user-denable
. The
capacitance
model
equation
is:
C
=
(C0)
+
(C1
2
F)
+
(C2
2
F
2
)
+
(C3
2
F
3
)
where
F
is
the
measurement
frequency
.
The
terms
in
the
equation
are
dened
with
the
specify
open
menu
as
follows:
N
N
N
N
N
N
N
N
C0
allows
you
to
enter
the
C0
term,
which
is
the
constant
term
of
the
cubic
polynomial
and
is
scaled
by
10
0
15
F
arads
.
N
N
N
N
N
N
N
N
C1
allows
you
to
enter
the
C1
term,
expressed
in
F/Hz
(F
arads/Hz)
and
scaled
by
10
0
27
.
N
N
N
N
N
N
N
N
C2
allows
you
to enter
the C2
term,
expressed
in
F/Hz
2
and
scaled
by
10
0
36
.
N
N
NN
NN
N
N
C3
allows
you
to
enter
the
C3
term,
expressed
in
F/Hz
3
and
scaled
by
10
0
45
.
N
N
N
N
N
N
N
N
NN
NN
NN
N
N
N
SHORT
denes
the
standard
type as
a short,
for calibrating
reection measurements
.
Shorts
are
assigned a
terminal
impedance
of
0
ohms
,
but
delay
and
loss
osets
may
still
be
added.
N
N
N
N
N
N
N
N
N
N
N
N
NN
LOAD
denes
the
standard
type
as
aload
(termination).
Loads
are
assigned
a
terminal
impedance equal to the system characteristic impedance Z0, but delay and loss osets
may still be added. If the load impedance is not Z0, use the arbitrary impedance standard
denition.
NNNNNNNNNN
NNNNNNN
FIXED
denes
the load as a xed (not sliding) load.
NNNN
NNNNNNNNNNNNNNNNNNN
SLIDING
denes the load as a sliding load. When such a load is measured during
calibration, the analyzer will prompt for several load positions
, and calculate the ideal load
value from it.
NNNNNNNNNNNNNNNNNNNN
OFFSET
denes the load as being oset.
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DELAY/THRU
denes the standard type as a transmission line of specied length, for
calibrating transmission measurements.
Application and Operation Concepts 6-79