N
NN
NN
NN
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
ARBITRARY
IMPEDANCE
denes
the standard
type to
be a
load, but
with an
arbitrary
impedance
(dierent
from
system
Z0).
N
NN
NN
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
NN
NN
NN
N
TERMINAL
IMPEDANCE
allows
you
to
specify
the
(arbitrary)
impedance
of
the
standard,
in
ohms
.
N
NN
NN
NN
NN
NN
NN
N
N
N
N
FIXED
denes the
load as
a xed
(not sliding)
load.
N
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
N
SLIDING
denes
the
load
as
a
sliding
load.
When
such
a
load
is
measured
during
calibration,
the
analyzer
will prompt
for several
load positions
,
and
calculate
the
ideal
load
value
from
it.
Any
standard
type
can be
further dened
with osets
in delay
,
loss
,
and
standard
impedance;
assigned
minimum
or
maximum frequencies
over which
the standard
applies; and
dened
as
coax
or
waveguide
.
The
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
NN
N
N
N
SPECIFY
OFFSET
softkey
provides
access
to
the specify
oset menu
(described
next).
The
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
LABEL STD
softkey allows
you
to
dene
a
distinct
label
for
each
standard,
so
that
the
analyzer
can
prompt
the
user
with explicit
standard labels
during
calibration
(such
as
SHORT
).
The
function
is
similar
to
dening
a
display
title
,
except that
the
label
is
limited
to
ten
characters
.
After
each
standard
is dened,
including osets
, the
N
N
N
N
N
N
N
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
NN
NN
NN
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
N
STD
DONE
(DEFINED)
softkey will
terminate the
standard
denition.
Specify
Oset
Menu
The
specify
oset
menu
allows
additional
specications
for
a
user-dened
standard.
F
eatures
specied
in
this
menu
are
common
to
all
ve
types
of
standards
.
Osets
may
be
specied
with
any
standard
type
.
This
means
dening
a
uniform
length
of
transmission
line
to
exist
between
the
standard
being
dened
and
the
actual
measurement
plane
.
(Example:
a
waveguide
short
circuit
terminator
,
oset
by
a
short
length
of
waveguide
.)
F
or
reection
standards
,
the
oset
is
assumed
to be
between the
measurement
plane
and
the
standard
(one-way
only).
F
or
transmission
standards
,
the
oset is
assumed to
exist
between
the
two
reference
planes
(in
eect,
the
oset
is the
thru). Three
characteristics of
the oset
can
be
dened:
its
delay
(length),
loss
,
and impedance
.
In
addition,
the
frequency
range
over
which
a
particular
standard is
valid can
be dened
with
a
minimum and
maximum frequency
. This
is particularly
important
for
a
waveguide
standard,
since
its behavior
changes rapidly
beyond its
cuto frequency
.
Note
that
several
band-limited
standards
can together
be dened
as the
same \class"
(see specify
class menu).
Then,
if
a
measurement
calibration
is
performed
over
a
frequency
range
exceeding
a
single
standard,
additional standards can be used for each portion of the range
.
Lastly,the
standard must be dened as either coaxial or waveguide
. If it is waveguide
,
dispersion eects are
calculated automatically and included in the standard model.
The following is a
description of the softkeys located within the specify oset menu:
NNNNNNNNNNNNNNNNNNNNNNNNN
NNNNNNNNNNNNN
OFFSET DELAY
allows you to specify the one-way electrical delay from the measurement
(reference) plane to the standard, in seconds (s). (In a transmission standard, oset delay is
the delay from plane to plane.) Delay can be calculated from the precise physical length of
the oset, the permittivity constant of the medium, and the speed of light.
In coax, group delay is considered constant. In waveguide, however, group delay is
dispersive,thatis, it changes signicantly as a function of frequency. Hence, for a waveguide
standard, oset delay must be dened as though it were a TEM wave (without dispersion).
6-80 Application and Operation Concepts