EasyManua.ls Logo

HP 8903B - Page 153

HP 8903B
240 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service Model 8903B
Total Change
Special in Oscilloscope
Function Vertical Gain
1.1
xl
1.5
+lO
1.8
+loo
1.11
+loo
Amplitude Limits (divisions pk-pk)
Minimum Maximum
4.8 5.2
3.0 3.3
1.2 1.3
4.9 5.1
Table
8B-6.
Amplitude Limits at A2TP5,
a
Step
7
Total Change
Special in Oscilloscope
Function Vertical Gain
1.11
xl
1.12
xl
1.13
+lo
1.14
+lo
Amplitude Limits (divisions pk-pk)
Special
Function Minimum Maximum
1.17
1.19 4.8
5.2
Amplitude Limits (divisions pk-pk)
Minimum Maximum
4.8
5.2
7.7 8.2
1.2 1.3
1.9 2.1
8.
On the Audio Analyzer, key in
1.11
SPCL. On the oscilloscope, set the gain back
to
xl.
(Check that the
waveform is still
5
divisions peak-to-peak.)
9. Connect the oscilloscope to A2TP4 (PGM AMP). On the Audio Analyzer, key in the Special Functions listed
in Table 8B-7.
For
each setting, decrease the oscilloscope’s gain by the amount indicated. The peak-to-peak
amplitude, displayed on the oscilloscope, should be within the limits indicated.
If
faulty, see Service Sheet
2
and
check the Programmable Gain Amplifier.
Table
8B-7.
Amplitude Limits at A2TP4,
a
Step
9
10. On the Audio Analyzer, key in
1.11
SPCL.
On
the oscilloscope, set the gain back to
xl.
(Check that the
waveform
is
still
5
divisions peak-to-peak.)
11.
Connect the oscilloscope to A2TP3 (OUT). The oscilloscope display should be between
4.9
and 5.1 divisions
peak-to-peak.
If
faulty, see Service Sheet
2
and check the Weighting Bandpass
and
High-Pass Filters.
12.
On the Audio Analyzer, key in AMPTD
300
mV. Decrease the vertical gain of the oscilloscope for a display
of
5
divisions peak-to-peak.
13. On the Audio Analyzer, key in the frequencies and the weighting bandpass
or
high-pass filter indicated in
Table
8B-8
for the
filters
installed in the instrument.
(If
no filter options are installed, go on to the next step.)
For
each setting, the peak-to-peak amplitude, displayed on the oscilloscope, should be within the limits indicated.
If
faulty, see Service Sheets 2, 2A, and 2B, and check the Weighting Bandpass and High-Pass Filters.
8B-8
rev.20JUN87

Related product manuals