206
Equivalent Front-Panel Key Commands
To Query the Increment Value.
Use the :INCRement? command to query the
increment value.
Syntax
:INCRement?
Example
OUTPUT 714;"RFG:FREQ:INCR?"
ENTER 714;Incr_value
This queries the increment value for the RF Gen Freq field.
NOTE: When querying a field setting or measurement result through HP-IB, the Test Set
always returns numeric values in HP-IB Units or Attribute Units, regardless of the
field’s current Display Units setting. Refer to "Attribute Units (AUNits)" on
page 73 and "HP-IB Units (UNITs)" on page 70.
To Set the Increment Mode.
Use the :INCRement:MODE commands to set the
increment mode to linear or logarithmic.
Syntax
:INCRement:MODE <LOGarithm or LINear>
Example
OUTPUT 714;"RFG:FREQ:INCR:MODE LOG"
This sets the increment mode for the RF Generator’s frequency to
logarithmic.
To Query the Increment Mode. Use the :INCRement:MODE? commands to
query the increment mode.
Syntax
:INCRement:MODE?
Example
ENTER 714;Mode$ ! returns LIN or LOG
OUTPUT 714;"RFG:FREQ:INCR:MODE?"
This queries the increment mode of the RF Generator’s frequency.