Step Function Range Input Signal
Adjustment
(limits)
Tolerance
(counts)
21
10 A
10 A 10.00 A
RV6 (±2) ±14
22
µA
400 µA380.0 µA @ 100 Hz
±62
23
4000 µA 3800 µA @ 100 Hz
±62
24
mA
40 mA 38.00 mA @ 100 Hz
±62
25 400 mA 380.0 mA @ 100 Hz
±62
26
10 A
10 A 10.00 A @ 100 Hz
±20
27
400
Short zero adjust
1
±1
28
380.0
±20
29
4 k 3.800 k
±20
30
40 k 38.00 k
±20
31
400 k 380.0 k
±20
32
4 M 3.800 M
±20
33
40 M 38.00 M
±40
34
400 0 to 100
Tone < 20
35 2 V 1.000 V
±12
36 Hz 4 9000 Hz @ 1 Vrms
±2
1
Perform zero adjustment using key.
Calibration Table