Step Function Range Input Signal
Adjustment
(limits)
Tolerance (counts)
972A 973A
29
mA
40 mA
38.00 mA @ 100 Hz
±61 ±61
30 38.00 mA @ 2 kHz
±61 ±61
31
400 mA
380.0 mA @ 100 Hz
±61 ±61
32 380.0 mA @ 2 kHz
±61 ±61
33
10 A
10 A
10.00 A @ 100 Hz
±19 ±19
34 10.00 A @ 2 kHz
±19 ±19
35
400
Short zero adjust
1
±1 ±1
36
380.0
±8 ±8
37
4 k 3.800 k
±8 ±8
38
40 k 38.00 k
±8 ±8
39
400 k 380.0 k
±8 ±8
40
4 M 3.800 M
±20 ±20
41
40 M 38.00 M
±40 ±40
42
400 0 to 100
Tone below approx 20
43 2 V 1.000 V
±12 ±12
44
100 µF90.0 µF9 (±2) ±29 ±29
45
10 µF9.00 µF10 (±2) ±12 ±12
46 10 nF Open zero adjust
1
±1 ±1
47 100 nF 90.0 nF
11 (±2) ±21 ±21
48 10 nF 9.00 nF
±21 ±21
49 1000 nF 900 nF
±12 ±12
50
1000 µF900 µF
±29 ±29
51 Hz (V ) 4 V 9000 Hz @ 1 Vrms
±2 ±2
52 Hz (A)
400 µA 9000 Hz @ 100 µA
±2 ±2
1
Perform zero adjustment using key.
Calibration Table
6 - 3