EasyManua.ls Logo

HP 974A - Resistance Measurement Specs; Continuity Test Specs; Diode Test Specs; Frequency Measurement Specs

HP 974A
27 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Resistance
Range Resolution Accuracy Test Current
Max
Open Circuit
Voltage
500 10 mΩ± (0.06% + 2)
1
< 800 µA
< 5.5 V
5.0 k 100 m
± (0.06% + 2) 50 k 1 < 80 µA
< 2.2 V
500 k 10 < 15 µA
5.0 M 100 ± (0.5% + 1) < 1.5 µA
50 M 1 k ± (1.0% + 2)
< 150 nA
1
After zero adjust of input leads. Zero adjust range up to 0.99.
Response time: 500 to 500 k — < 2 seconds, 5 M to 50 M — < 10 seconds.
Continuity
Measurement Current: 0.8 mA maximum Open circuit voltage: < 5.5 Vpeak
Displayed resistance: 0 to 499.99 Input protection: 500 Vrms (sinewave)
Alarm: Tone when input < 100 Ω ± 50 Resolution: 10 m (<100 mSec response time)
Diode
Measurement current: +1.0 mA nominal @ 0.6 V Open circuit voltage: < 5.5 Vpeak
Displayed Voltage: 0 V to 4.999 V Input protection: 500 Vrms (sinewave)
Accuracy: ± (1% + 2) Resolution: 100 µV
Frequency (AC Coupled)
Frequency Range Resolution Accuracy Input Voltage (rms)
10 Hz to 99.99 Hz 0.01 Hz
± (0.05% + 1)
0.45 mV to 500 V90 Hz to 999.0 Hz 0.1 Hz
900 Hz to 9.999 Hz 1 Hz
9.00 kHz to 99.99 kHz 10 Hz .7 V to 100 V
90 kHz to 200 kHz 100 Hz 1.5 V to 100 V
Specifications
1-19