EasyManua.ls Logo

HP 974A - Page 23

HP 974A
27 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Step Function Range Input Signal
Adjustment
(limit)
Tolerance
(counts)
17
V
50 V
48.000 V @ 10 kHz
C3 (±20) ±270
18 48.000 V @ 100 kHz
±1740
19 48.000 V @ 200 Hz
±270
20 500 mV 480.00 mV @ 10 kHz
±366
21 750 V 750.0 V @ 200 Hz
±105
22
500
Short zero adjust
1
±1
23
480.00 9 (±5) ±30
24
5 k 4.8000 k
±30
25
50 k 48.000 k
±30
26
500 k 480.00 k
±30
27
5 M 4.8000 M
±242
28
50 M 48.000 M
±482
29
500 µA500 µA
Short
±2
30
480.00 µA
±146
31
50 mA
50 mA 48.000 mA
±146
32
500 mA
40 mA 480.00 mA
±146
33
10 A
10 A 10.000 A
10 (±10) ±72
34
500 µA500 µA 480.00 µA @ 200 Hz
±500
35
50 mA
50 mA 48.000 mA @ 200 Hz
±500
36
500 mA
500 mA 480.00 mA @ 200 Hz
±500
37
10 A
10 A 10.00 A @ 200 Hz
±120
38
500 0 to 150
Tone < 100
39 5 V 1.000 V
±102
40 Hz 5 V 9.000 kHz @ 1 Vrms
±5
1
Perform zero adjustment using key.
Calibration Table