GEO416 - GEO416GS
EN - 22
9.1.1. Reference standards
Safety: IEC/EN61010-1, IEC/EN61557-1, IEC/EN61557-5
Measuring accessories’ safety: IEC/EN61010-031
Technical literature: IEC/EN61187
Insulation: double insulation
Mechanical protection: IP50 according IEC/EN60529
Pollution level: 2
Measurement category: CAT III 240V (to earth), maximum 415V between inputs
Max height of use: 2000m (6562ft)
9.1.2. General features
Mechanical features
Dimensions (L x W x H): 235 x 165 x 75mm (9 x 6 x 3in)
Weight (including batteries): 1kg (2lv)
Power supply
Battery type: 6x1.5V batteries type AA LR6 or
6x1.2V rechargeable batteries NI-MH AA LR6 2100mA
Low battery indication: symbol “ ” is displayed
Battery life: approx. 500 tests
Auto Power OFF: after 3 min of idleness
Display
Features: LCD custom with back-light 73x65 mm
Memory
Features: 999 memory locations
PC connection
Features: optoinsulated port for bi-directional communication
9.2. ENVIRONMENT
9.2.1. Operating environmental conditions
Reference calibration temperature: 23 ± 5°C (73 ± 41°F)
Working temperature: 0 ÷ 40°C (32°F ÷ 104°F)
Maximum relative humidity: <80%RH
Storage temperature: -10 ÷ 60°C (14°F ÷ 140°F)
Storage humidity: <80%RH
This instrument complies with the requirements of the European Low Voltage
Directive 2006/95/CE (LVD) and EMC Directive 2004/108/CE
9.3. ACCESSORIES
9.3.1. Standard and optional accessories GEO416
See enclosed packing list
9.3.2. Standard accessories GEO416GS
Set of 4 cables, banana-banana, L=1m KIT416CV
Set of 4 alligator clips COC4-UK
Carrying bag BORSA2000N
ISO9000 calibration certificate
User manual