EasyManua.ls Logo

HT HT7051 - Technical Specifications; DC Voltage Specifications; AC TRMS Voltage Specifications; Insulation Resistance Intrinsic Uncertainty

HT HT7051
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HT7051 - HT7051N
EN - 28
10 TECHNICAL SPECIFICATIONS
Accuracy is calculated as [% rdg + (dgt number * resolution)] at 23 ± 5°C,< 80%RH
DC VOLTAGE
Range Resolution Accuracy Protection against overloads
10 600V
1V
(2.0rdg+2 dgt)
CAT IV 600V to earth
AC TRMS VOLTAGE
Range Resolution Accuracy Protection against overloads
10 600V
1V
(2.0rdg+2dgt)
CAT IV 600V to earth
INSULATION RESISTANCE – INTRINSIC UNCERTAINTY (AT THE REFERENCE CONDITIONS)
Range Test voltage Resolution Accuracy (*)
0.01M 0.19M (**) 100V
1%rdg
(5rdg.+7dgt)
0.20M 199G 100V
(5rdg+3dgt) if
nA
VoltageTest
Rmis
5
(20rdg+3dgt) if
nA
VoltageTest
Rmis
5
0.20M 499G 250V
0.20M 999G 500V
0.20M 1.99T 1000V
0.20M 4.99T 2500V
0.20M 9.99T
5000V
(*) Load Capacitive < 1nF
(**) Outside range indicated by IEC/EN61557
INSULATION RESISTANCE– OPERATING UNCERTAINTY (COMPLIANCE WITH IEC/EN61557-1)
Range Test voltage Resolution Uncertainty (*)
0.01M 0.19M (**) 100V
1% rdg
(12rdg+7dgt)
0.20M 199G 100V
(12rdg+3dgt) if
nA
VoltageTest
Rmis
5
(20rdg+3dgt) if
nA
VoltageTest
Rmis
5
0.20M 499G 250V
0.20M 999G 500V
0.20M 1.99T 1000V
0.20M 4.99T 2500V
0.20M 9.99T
5000V
(*) Load Capacitive < 1nF
(**) Outside range indicated by IEC/EN61557
GENERATED VOLTAGE (*)
Test mode Nominal test voltages Accuracy
FIX 100V, 250V, 500V, 1kV, 2.5kV, 5kV
-0%, +10% + 15V
ADJUSTABLE
100 1000V, step 25V
1000 5000V, step 50V
RAMP
100 1000V, step 25V
1000 5000V, step 50V
(*) Test voltage compliance with IEC /EN 61557-2
TEST CURRENT
Test voltage
Test current
100 5000V 1mA Test current 3mA (*)
(*) Automatically controlled test current.
TEST TIME
Setting range
Resolution
5s – 99min 59s 1s