10 TECHNICAL SPECIFICATIONS
Accuracy is calculated as [% rdg + (dgt number * resolution)] at 23 ± 5°C,< 80%RH
DC VOLTAGE
Range Resolution Accuracy Protection against overloads
10 600V
1V
(2.0rdg+2 dgt)
CAT IV 600V to earth
AC TRMS VOLTAGE
Range Resolution Accuracy Protection against overloads
10 600V
1V
(2.0rdg+2dgt)
CAT IV 600V to earth
INSULATION RESISTANCE – INTRINSIC UNCERTAINTY (AT THE REFERENCE CONDITIONS)
Range Test voltage Resolution Accuracy (*)
0.01M 0.19M (**) 100V
1%rdg
(5rdg.+7dgt)
0.20M 199G 100V
(5rdg+3dgt) if
nA
VoltageTest
Rmis
5
(20rdg+3dgt) if
nA
VoltageTest
Rmis
5
0.20M 499G 250V
0.20M 999G 500V
0.20M 1.99T 1000V
0.20M 4.99T 2500V
0.20M 9.99T
5000V
(*) Load Capacitive < 1nF
(**) Outside range indicated by IEC/EN61557
INSULATION RESISTANCE– OPERATING UNCERTAINTY (COMPLIANCE WITH IEC/EN61557-1)
Range Test voltage Resolution Uncertainty (*)
0.01M 0.19M (**) 100V
1% rdg
(12rdg+7dgt)
0.20M 199G 100V
(12rdg+3dgt) if
nA
VoltageTest
Rmis
5
(20rdg+3dgt) if
nA
VoltageTest
Rmis
5
0.20M 499G 250V
0.20M 999G 500V
0.20M 1.99T 1000V
0.20M 4.99T 2500V
0.20M 9.99T
5000V
(*) Load Capacitive < 1nF
(**) Outside range indicated by IEC/EN61557
GENERATED VOLTAGE (*)
Test mode Nominal test voltages Accuracy
FIX 100V, 250V, 500V, 1kV, 2.5kV, 5kV
-0%, +10% + 15V
ADJUSTABLE
100 1000V, step 25V
1000 5000V, step 50V
RAMP
100 1000V, step 25V
1000 5000V, step 50V
(*) Test voltage compliance with IEC /EN 61557-2
TEST CURRENT
Test voltage
Test current
100 5000V 1mA Test current 3mA (*)
(*) Automatically controlled test current.
TEST TIME
Setting range
Resolution
5s – 99min 59s 1s