EasyManua.ls Logo

HT HT7051 - Generated Voltage Specifications; Test Current Specifications; Test Time Specifications; Capacitance Measurement Specs

HT HT7051
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HT7051 - HT7051N
EN - 29
CAPACITANCE MEASUREMENT
Range Resolution Load resistance Test voltage (Vn) Accuracy
1nF 999nF
1nF
5M
Vn 5000V
(10rdg+5dgt)
1.00
F 5.00
F 0.01
F
1nF 999nF
1nF
Vn 2500V
1.00
F 9.99
F 0.01
F
10.0
F 19.9
F 0.1
F
1nF 999nF
1nF
Vn 1000V
1.00
F 9.99
F 0.01
F
10.0
F 49.9
F 0.1
F
Charge time of capacitance (0V 5000V): < 3s x 1F
Discharge time of capacitance (5000V 25V): < 5s x 1F
LEAKAGE CURRENT
Range
Resolution Accuracy
0.1nA 99.9nA
0.1nA
(7rdg+3dgt) if
nA
VoltageTest
Rmis
5
(22rdg+3dgt) if
nA
VoltageTest
Rmis
5
100nA 999nA
1nA
1.00
A 9.99
A 0.01
A
10.0
A 9.99
A 0.1
A
100
A 999
A 1
A
1.00mA 2.5mA
0.01mA
PI – DAR
Range Resolution Accuracy (*)
0.01 9.99
0.01
(5rdg+3dgt) if
nA
VoltageTest
Rmis
5
(20rdg+3dgt) if
nA
VoltageTest
Rmis
5
(*) Load capacitance < 1nF
10.1 REFERENCE GUIDELINES
Instrument safety: IEC/EN61010-1, IEC/EN61557-1, IEC/EN61557-2
Measuring accessory safety: IEC/EN 61010-031:2002 + A1:2008
EMC: IEC/EN61326-1
Technical documentation: IEC/EN61187
Insulation: double insulation
Protection: IP40 (open case), IP53 (closed case) according to IEC
/EN60529
Polluting level: 2
Measurement category: CAT IV 600V (to earth), max 600V between inputs
Maximum height of use: 2000m (6561ft)
10.2 GENERAL CHARACTERISTICS
Mechanical characteristics
Dimensions (L x W x H): 360 x 310 x 195mm (14 x 12 x 8in)
Weight: 3.5kg (7.8lb)
Power supply
External power supply: 220-240VAC, 50-60Hz, 20VA (HT7051)
110-120VAC, 50-60Hz, 20VA (HT7051N)
Internal battery type: 1x4.8V, 3800mAh NiMH rechargeable by mains
Recharging time: 4 hours
Fuse: T 200mA H 250V, Ir:1,5kA
Low battery indication: symbol “ ” at display.