EasyManua.ls Logo

Humboldt Digital MasterLoader HM-3000.3F - Page 23

Humboldt Digital MasterLoader HM-3000.3F
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
To assign a test to the selected channel, press the corresponding function
keytothetestyouwishtoselect.Pressingakeywilltogglebetween“Y”
and“N”.Forexample,press to choose Marshall Test, which will
changethe“N”toa“Y”nexttoMarshall.Thiswouldindicatethatthe
channel selected would be set up to run Marshall Test.
NOTE: Multiple tests can be assigned to a single channel, however, multiple
channels of the same category cannot be assigned to the same test.
Example: Ifchannel1isconguredwitha10,000lb(50KN)loadcell,and
channel2isconguredwitha5,000lb(25KN)loadcell,itispossibletouse
either channel to perform a Marshall test. However, only one of the channels
can be used as the load channel (Stability channel).
To return to any prior screen you can always press
Press (ENGINEERINGSETUP).Youshouldseethisscreen:
Press (CONFIGURATION)Youwillbeaskedforapassword,enter:
You should see this screen:
Press (SELECTCHANNEL)toscrollthroughandchoosethechannel
you wish to configure test for. Once you have selected a channel press
(CONFIGURETEST).Youshouldseethisscreen:
* CONFIGURATION
* CALIBRATION
* LOGGING INTERVAL
CONFIGURATION screen
* SELECT CHANNEL “CH1”
* CONFIGURE CHANNEL
* CONFIGURE TEST
* CBR/LBR TEST N
* MARSHALL TEST N
* USER DEFINED N

Related product manuals