4433
Probe Functional Test Specification
A
B
Probe
Testing
Sites
A. OP2HS,
OP3HS,
D920-PP,
D930-PP,
FD1-PP
and
FD3-PP
tests
on
forearm.
B. VP
4HS,
VP5HS,
VP8HS,
VP10HS
and
EZ8
tests
on
arm
(brachial
artery).
D9220-PP,
D9330-PP,
FD1-PP,
Switch the unit on. Test the probe for noise and signal levels
FD33-PP
referring to the figure above for a suitable test site, (using a
standard probe as a reference).
OP22HS,
OP33HS
These must be tested with a MD2-P or FD2-P.
Check that the unit defaults to Fetal Mode, and that the frequency
indication is correct. Test audio performance as above.
8.3 Functional Test - Obstetric Probes OP2HS, OP3HS, D920-P, D930-P,
FD1-P, FD3-P