36 SLIT LAMP HS-5000 5X------------------- ----------------------------------------------------------------
11.3. Guidance and manufacturer’s declaration – electromagnetic
immunity
The EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an
environment.
Immunity
test
IEC
60601
test
level
Complianc
e
level
Electromagnetic environment - guidance
Conducted
RF
IEC
61000-4-6
Radiated
RF
IEC
61000-4-3
3 Vrms
150
kHz to
80MHz
3 V/m
80
MHz to
2.5GHz
3 Vrms
150 kHz to
80MHz
3 V/m
80MHz to
2.5GHz
Portable and mobile RF communications
equipment should be used no closer to any
part of the EUT, including cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
w
here P is the maximum output power
rating of the transmitter in watts (W)