40 SLIT LAMP HS-5000 5X------------------- ----------------------------------------------------------------
11.6. Guidance and manufacturers declaration – electromagnetic
immunity
The EUT is intended for use in the electromagnetic environment specified below.
The customer or the user of the EUT should assure that it is used in such an
electromagnetic environment.
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Conducted
RF
IEC
61000-4-6
Radiated
RF
IEC
61000-4-3
3 Vrms
150 kHz to
80MHz
3 V/m
80 MHz to
2.5GHz
3 Vrms
150 kHz to
80MHz
3 V/m
80MHz to
2.5GHz
The EUT must be used only in a
shielded location with a minimum RF
shielding effectiveness and, for each
cable that enters the shielded
location with a minimum RF shielding
effectiveness and, for each cable that
enters the shielded location
Field strengths outside the shielded
location from fixed RF transmitters,
as determined by an
electromagnetic site survey, should
be less than 3V/m.