-------------------------------------------------------------------------------------SLIT LAMP HS-5000 5X 41
NOTE 1) These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.
NOTE 2) It is essential that the actual shielding effectiveness and filter
attenuation of the shielded location be verified to assure that they meet the
minimum specification.
a
Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an
electromagnetic site survey should be considered. If the measured field strength
outside the shielded location in which the EUT is used exceeds 3V/m, the EUT
should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary,
such as relocating the EUT or using a shielded location with a higher RF
shielding effectiveness and filter attenuation.