EasyManua.ls Logo

Hydronix Hydro-Probe - Figure 22: Example of a Good Brix Calibration; Figure 23: Example of a Bad Brix Calibration

Default Icon
53 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sensor Integration and Material Calibration Chapter 3
Configuration and Calibration Guide HD0679 Rev 1.2.0 39
5. The average of the laboratory results and the Filtered Unscaled value make up one
calibration point.
6. Steps 3-5 should be repeated for additional calibration points. Ideally calibration points
should be collected to cover the entire expected Brix range of the material.
Hydro-Com software should be used to calculate the calibration coefficients and to update the
sensor with the calibration.
9.1 Good/Bad Brix Calibration
A good Brix calibration is achieved by analysing the material over the working range. A good
spread of points is necessary to provide higher accuracy.
Figure 22 shows a good calibration with all points close to the best fit curve.
Improved Calibration
Current Calibration
Brix
Unscaled
Figure 22: Example of a Good Brix Calibration
Figure 23 is an example of a bad Brix calibration, this is evident as the points are not all close
to the best fit curve.
Unscaled
Brix
Improved Calibration
Current Calibration
Figure 23: Example of a Bad Brix Calibration
For full details on the use of Hydro-Com see user guide HD0682.

Table of Contents