GENERAL EQUIPMENT CARE
SONOPULSE III is a protective CLASS II device with applied
part type BF of safety and protection. SONOPULSE III
works in power supply tension in the range of 100 - 240V ~
50/60 Hz. Just connect the device to the power line and it
will perform the selection of power tension automatically. The
connector cable to the power supply line is detachable. The
device uses the power line plug as a resource to electrically
separate its circuits in relation to the power supply line in all
poles.
ELECTRICAL FEED
In the rear part of SONOPULSE III there is a protection fuse.
To replace it, turn the device o the power supply line
and with the help of a screwdriver, remove the protection lid,
disconnect the fuse, perform the replacement and reinsert
the lid.
Always use the fuses indicated by IBRAMED. Use a fuse for
nominal current of 5.0A, operation tension of 250V~ and
snap action model 20AG (50A rupture current).
SONOPULSE III does not need any type of power stabilizer.
Never use a power stabilizer.
NOTES
Before turning on SONOPULSE III make sure:
The tension and frequency of the local power supply line of
the establishment where the device is installed are equal to
the one described on the label describing characteristics of
tension and power located at the rear part of the device.
To prevent electrical shock, do not use the plug in the
device with extension cables, or any other types of sockets
except the terminals connect perfectly in the receptacle.
Cleansing and disinfection must always be performed with
the power plug o of the power supply line.
Maintenance and technical assistance of SONOPULSE III
must always be performed at unauthorized service, only by
qualied technicians.
Inside the device there are dangerous tensions.
Never open the device.