5 - 6
5-3 MAIN-B UNIT ADJUSTMENT (PCR2500 only; continued)
ADJUSTMENT OPERATION VALUE
S-METER
[NS0]–[WS6]
1 Set the SSG as;
Frequency : Specifi ed frequency*
Modulation : None
Level : Specifi ed level
Automatic adjustment
2 Push [SET•SKIP] to store the adjustment value.
3 Repeat 1–2 for each specifi ed frequency and value for [NS0] to [WS6].
S-METER FLATNESS
[L5L]–[L8H]
1 Set the SSG as;
Frequency : Specifi ed frequency*
Modulation : None
Level : Specifi ed level
Automatic adjustment
2 Push [SET•SKIP] to store the adjustment value.
3 Repeat 1-2 for each specifi ed frequency and value for [L5L] to [L8H].
CENTER METER
[CML] (LOW)
1 Set the SSG as;
Frequency : 3 kHz lower than specifi ed frequency*
Modulation : None
Level : +34 dBµ (–73 dBm)
Automatic adjustment
2 Push [SET•SKIP] to store the adjustment value.
[CMH] (HIGH) 1 Set the SSG as;
Frequency : 3 kHz higher than specifi ed frequency*
Modulation : None
Level : +34 dBµ (–73 dBm)
2 Push [SET•SKIP] to store the adjustment value.
SQUELCH
[SQL]
1 Set the SSG as;
Frequency : Specifi ed frequency*
Mode : FM
Modulation : 1 kHz
Deviation : 3.5 kHz
Level : –13 dBµ (–120 dBm)
Automatic adjustment
2 Set the [SQL] value to close the squelch with [DIAL]. Then set the
[SQL] value at the point where the audio signals just appear.
3 Turn the SSG output OFF, and verify that the squelch is closed.
4 Push [SET•SKIP] to store the adjustment value.
RSSI
[RS0]–[RS6]
1 Set the SSG as;
Frequency : Specifi ed frequency*
Modulation : None
Level : Specifi ed level
Automatic adjustment
2 Push [SET•SKIP] to store the adjustment value.
3 Repeat 1–2 for each specifi ed frequency and value for [RS0] to [RS6].
RSSI FLATNESS
[R5L]–[R8H]
1 Set the SSG as;
Frequency : Specifi ed frequency*
Modulation : None
Level : Specifi ed level
Automatic adjustment
2 Push [SET•SKIP] to store the adjustment value.
3 Repeat 1–2 for each specifi ed frequency and value for [R5L] to [R8H].
*Displayed on the controller's display.
The output level of the standard signal generator (SSG) is indicated as the SSG's open circuit.
Refer to the “ADJUSTMENT ITEM LIST“ on page 5-7.