C-15
SNAP* Heartworm RT
Test Description
The SNAP Heartworm RT Test is an in vitro diagnostic test for the semiquantitative detection of
Dirofilaria immitis (D. immitis) antigen in canine and feline whole blood, serum, or plasma.
IMPORTANT: SNAPshot Dx supports the SNAP Heartworm RT Test with canine samples only—tests
with feline samples must be read visually.
Note:
Any color development in the sample spots on a SNAP Heartworm RT indicates a positive test
result.
Preparing and Running the SNAP Heartworm RT Test
Important: Allow all SNAP Heartworm RT test components to equilibrate at room temperature for at
least 30 minutes. Do not heat.
1. Enter the patient information on the IDEXX VetLab Station (for more information, see the
Analyzing Samples” chapter of the IDEXX VetLab Station Operator’s Guide).
2. Once the patient’s name appears in the Pending list on the SNAPshot Dx Home screen, tap
the patient’s name and then tap Select. The Identify Sample screen appears.
3. Select Heartworm RT as the SNAP Test Type.
4. Select the applicable sample type.
5. Tap Optional Prep Instructions if you would like to view on-screen sample preparation
instructions.
a. Hold the provided pipette vertically and dispense 3 drops of sample into a disposable
sample tube.
b. Holding the bottle vertically, add 4 drops of conjugate to the sample tube.
c. Cap the sample tube and invert it 4–5 times to mix.
6. Tap Next and follow the on-screen instructions to properly prepare the SNAP test:
a. Place the device on a horizontal surface.
b. Add the entire contents of the sample tube to the sample well, being careful not to
splash the contents outside of the sample well. The sample will flow across the result
window, reaching the activation circle in 30–60 seconds. Some sample may remain in
the sample well.
c. When color FIRST appears in the activation circle, press the activator firmly until it is
flush with the device body.
Note: Some samples may not flow to the activation circle within 60 seconds. In this case,
press the activator after the sample has flowed across the result window.
7. Insert the SNAP device into the SNAPshot Dx Analyzer immediately. The SNAPshot Dx
Analyzer automatically begins to process the patient sample and the Home screen displays.
The patient’s name moves from the Pending list to the In Process list.
When analysis is complete, the patient’s name will move from the In Process list to the Results
list (you may also receive a “New Results” alert message and/or hear a beep on the IDEXX
VetLab Station, depending on the option(s) you selected in the New Results Alert tab on the
IDEXX VetLab Station’s Settings screen). You can then remove the SNAP device from the
analyzer by gently pushing the SNAP device until you hear a click. The device will slide out of
the port.
SNAP Test Procedures

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Idexx SNAPshot Dx and is the answer not in the manual?

Idexx SNAPshot Dx Specifications

General IconGeneral
BrandIdexx
ModelSNAPshot Dx
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals