User Manual ToFscan 25
• Stimulation or ECG electrodes:
- Capable of supporting up to 300 volts with 60 mA current.
- Contact surface must be greater than 1.8cm² (0,28in²).
Examples of recommended electrodes:
- RED DOT electrodes ref.2560 from the company 3M
- F9047 electrodes from the company FIAB
Please check with your authorized distributor or with the manufacturer which products are available
and approved in your country.
Data transfer
• Optical output for ber optic connection
• TOF-RS1 and TOF-RS2 are the only recommended Optic-Serial (RS232) cable to connect ToFscan
to other monitors.
Power supply
• 2900 mAh (minimum)/ 3.7V Lithium-Ion battery (comes with thermal protection and protection
against short-circuits)
• Battery power for about one month with normal use (10 TOF measurements a day).
• Charger / External power supply (continuous 5V, 1 A minimum).
Dimensions / Weight
• 60x150x55 mm (monitor only).
• 320 g (approximately) with battery and cable and electrode. (190g excluding cable).
Warranty
• Length of warranty: 2 years, 6 months for accessories and sensor.
Power consumption
• With power supply connected: 1 Watt while in use and 0,1 Watt while in sleep mode.
Environment
Shipping and storage conditions
The ToFscan and its accessories must be stored or transported under the following
restrictions and conditions. These conditions apply to non-operational storage and transport
conditions.
Temperature 10°C to +50°C
Humidity 15% to 95% (without condensation)
Pressure 500 hPa to 1060 hPa
For storage and transport, the original packaging must be used.
Protect the ToFscan from pikes in temperature which can cause condensation.
Operating environment
Reminders:
Risk of explosion: do not use the ToFscan in a ammable environment or in locations where
ammable anaesthetics may be accumulated.
The ToFscan is not designed to operate in an environment where there are SCANNERS, M.R.I.’s or any
other devices creating large magnetic elds. In order to limit electrostatic discharge, the humidity
must be maintaining above 35% and an antistatic oor is recommended.
The ToFscan is designed to operate safely under the following conditions. Situations other than those
described are likely to compromise the reliability of the device.
Temperature 10°C to +40°C
Humidity 35% to 90% (without condensation)
Pressure 700 hPa to 1060 hPa