User Manual ToFscan 6
The output from electrical stimulation causes nociceptive stimulations and the intensity of
these stimulations must be adapted to the patient’s analgesic level.
Do not use the ToFscan on patients wearing pacemakers without verifying and identifying the
possible consequences. The user must take all the necessary precautions during the operation
for this kind of patient.
Prior to use, check that no other equipment, device or material is in contact with the electrodes.
Sensors and electrodes should only be in contact with clean and healthy skin
Prior to each use, check that the device, the display and the cables (electrodes and sensors) are
not damaged. Do not use if a part is damaged
Handle the device with care to prevent any fall.
The ToFscan should be used during a limited time on one patient at a time and cleaned between
patients.
The ToFscan can be used fully or partially during the surgery. The duration should not exceed
24H.
After positioning one of the sensors in contact with the patient, check regularly at least every 2 to
3 hours that the sensor does not cause excessive pressure or stress on the patient’s skin. If skin
appearance changes, change the site sensor.
Portable RF communication equipment (including peripherals such as antenna cables and
external antennas) should not be used within 30 cm (1 foot) of any part of the ToFscan, including
accessories specied by the manufacturer.
Use of this equipment adjacent to or stacked with other equipment should be avoided because
it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally
Use of accessories, transducers and cables other than those specied or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
Use cables other than the one recommended by the manufacturer of this equipment could
result in increased cyber security risks.
Caution
Read this entire manual carefully before using the ToFscan.
Never put the ToFscan or any of its parts or accessories into an autoclave.
Do not submerge or spray the appliance or any of its parts with liquids.
The ToFscan and its parts are not suitable for the processes of gas, radiation (gamma or other),
water bath, steam or heat sterilisation.
Follow the instructions given in the “Cleaning and disinfection” section for cleaning and disinfecting
the ToFscan.
The ToFscan carries an internal lithium-ion battery. The ToFscan battery should under no
circumstances be dissembled, modied or replaced. Any tampering with the battery poses the risk

Other manuals for Idmed ToFscan