EasyManua.ls Logo

IFM Electronic efector300 SI5000 - Setup and Settings for Water; Change the Switch Point (Optional)

IFM Electronic efector300 SI5000
13 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8
6 Set-up and settings for water
(For media other than water → 7�1: Low flow adjustment)�
Switch on the supply voltage�
> All LEDs light and go out again step by step� During this time the output is
closed (if configured as normally open)� The unit is in the operating mode�
Let the normal flow circulate in the installation�
Check the display and determine further actions�
1


The factory setting is suitable for the applica-
tion�
No further settings are required�
2


Your normal flow is below the representation
range of the display
2 setting options:
Change the switch point (→ 6�1)�
Carry out high flow adjustment (→ 6�2)�
3


Your normal flow exceeds the representation
range of the display (LED 9 flashes)�
Carry out high flow adjustment (→ 6�2)�
You can restore the factory setting any time� (→ 7�3)�
6.1 Change the switch point (optional)
For the factory setting the switch point is at LED 7� A change makes sense if:
the display shows example 2�
the flow fluctuates much or pulsates�
if a faster response time of the unit is requested (low switch point = fast re-
sponse with rising flow, high switch point = fast response with falling flow)�
Briefly press the pushbutton
or
> The switch point LED flashes�
Press the pushbutton
or as often as required� Each press of the push-
button shifts the LED by one position in the indicated direction�
Note: If no pushbutton is pressed for 2 s, the unit returns to the operating mode
with the newly set value�

Related product manuals