EasyManua.ls Logo

IFM Electronic TR7432 - Maintenance and Service; Min;Max Temperature Logging; Factory Reset Procedure

IFM Electronic TR7432
17 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
13
UK
9.4.2 Zero point calibration
Select [COF] and enter value. Setting range:
-10.0...10.0 °C in steps of 0.1 °C.
-18.0 ...18.0 °F in steps of 0.1° F.
The internal measured value "0" is shifted by this value.
9.4.3 Setting the switching delays
[dS1] / [dS2] = switch-on delay for OUT1 / OUT2.
[dr1] / [dr2] = switch-off delay for OUT1 / OUT2.
Select [dS1], [dS2], [dr1] or [dr2] and set a value between 0.1 and 50 s
(at 0.0 the delay time is not active).
9.4.4 Setting of the error behaviour of the outputs
Select [FOU1] or [FOU2] and set the value:
[On] = Output 1 / 2 switches ON in case of a fault. -
[OFF] = Output 1 / 2 switches OFF in case of a fault. -
9.4.5 Setting of the switching logic for OUT1
Select [P-n] and set [PnP] or [nPn].
9.5 Service functions
9.5.1 Reading of the min/max values for the system temperature
Select [HI] or [LO], briefly press [Set].
[HI] = maximum value, [LO] = minimum value.
Delete memory:
Select [HI] or [LO].
Press [Set] and keep it pressed until [----] is displayed.
Press [Mode/Enter] briefly.
9.5.2 Reset of all parameters to factory setting
Select [rES].
Press [Set] and keep it pressed until [----] is displayed.
Press [Mode/Enter] briefly.
We recommend taking down your own settings in the table before carrying
outareset(→13Factorysetting).

Related product manuals