RF
GENERATE
EXTERNAL MODULATION
CALIBRATION
PREREQUISITES:
1-8-1
POWER
SUPPLY CALIBRATION
EQUIPMENT REQUIRED:
FIG
URES:
1-8.7 RF
GENERATE
FM
DEVIATION
CALIBRATION
1.8.8 RF
GENERATE PM DEV!ATION
CALIBRATION
1-8-9 RF
GENERATE
AM MODULATION
CALIBRATION
1
AUDIO ANALYZER
1
MODULATION
ANALYZER
1-5
AUDIO ANALYZER
coM-1208
(18) (12)
MODUI-ATlON
ANALYZER
HlGH IMPEDANCE
OUTPUT
00603002
Figure 1-5 External
Modulation
Calibration
Setup
STEP PROCEDURE
1. Apply Power
to
COM-12O8.
Allow
5
minute warm-up
period.
2.
Connect Test Equipment
as shown
in Figure
1-5.
3.
Press
GEN
Test Mode
Key
(3)
to access
RF
Generate Operation Screen.
4.
Set
RF Field
to
101.0000 MHz.
5.
Set Output to
T/R
and Output Level
to
-20
dBm.
6. Set
All
Modulation
Sources to OFF.
7.
Set
Modulation
Analyzer for 3 kHz Low-Pass
Filter,
300
Hz High-Pass Filter,
Peak+
and
AM
Measu rement.
Set
EXT Modulation Type tor
AM.
Set
Audio
Analyzer for 1 kHz
tone at approximately
5.657
Vrms
output.
Adjust
Audio
Analyzer
for Modulation Analyzer
reading
of 80%.
8.
9.
1-25
1-8-10
RF
GENERATE
EXTERNAL
MODULATION
CALIBRATION
PREREQUISITES:
1-8-1
POWER
SUPPLY
CALIBRATION
1-8-7
RF
GENERATE
FM
DEVIATION
CALIBRATION
1-8-8
RF
GENERATE
PM
DEVIATION
CALIBRATION
1-8-9
RF
GENERATE
AM
MODULATION
CALIBRATION
EQUIPMENT
REQUIRED:
FIGURES:
AUDIO ANALYZER
HIGH IMPEDANCE
OUTPUT
1-5
AUDIO
ANALYZER
MODULATION
ANALYZER
COM-120B
(18) (12)
EXT
MOD
T/R
MODULATION
ANALYZER
INPUT
00603002
Figure
1-5
External
Modulation
Calibration
Setup
STEP
PROCEDURE
1.
Apply
Power
to
COM-120B.
Allow
5
minute
warm-up
period.
2.
Connect
Test
Equipment
as
shown
in
Figure
1-5.
3.
Press
GEN
Test
Mode
Key
(3)
to
access
RF
Generate
Operation
Screen.
4.
Set
RF
Field
to
101.0000
MHz.
5.
Set
Output
to
TIR
and
Output
Level
to
-20
dBm.
6.
Set
All
Modulation
Sources
to
OFF.
7.
Set
Modulation
Analyzer
for
3
kHz
Low-Pass
Filter,
300
Hz
High-Pass
Filter,
Peak+
and
AM
Measurement.
8.
Set
EXT
Modu
lation
Type
for
AM.
9.
Set
Audio
Analyzer
for
1
kHz
tone
at
approximately
5.657
Vrms
output.
Adjust
Audio
Analyzer
for
Modulation
Analyzer
reading
of
80%.
1-25
Scans by ArtekMedia © 2008