Probe malfunctions:
If the probe electronics report a malfunction, this will also be displayed.
Fault in the probe
hardware
e.g. broken cable, hardware
fault
0% or 100% not calibrated or
wrong way round; calibrated
measuring range too small
Probe temperature too
high, occurs with
T
probe
≥105°C
Ambient temperature of probe
probably too high
Heat insulation measures on
the flange must not be
carried out on the electronic
part
e.g. broken cable, connections
wrong way round
Every error code (except “8.8.8”) causes the extra relay to be switched off.
The 4 mA .. 20 mA output then jumps to 0 mA within under 3 s
The flashing code „120“ is active if the level limit is passed. This is not
considered an error code. The 4 mA .. 20 mA output stays at its level
equivalent value.