Tempering flammable liquids:
For tempering flammable liquids, an external On/Off
temperature
con-
troller (not included with the magnetic stirrer) is required and must be
installed. The controller switches off the magnetic stirrer before the
media reaches the ignition point.
Safety feature:
The unit has an additional internal temperature sensor which will shut
the heating function off, in the event of a temperature control mal-
function.
Switch off the heat control knob, allow the unit to cool down, and then
switch the heat control knob on again.
Cleaning
Unplug the magnetic stirrer.
Clean carefully with a damp cloth and warm water.
Do not clean with acetone or other organic liquids, except
alcohol.
In extreme cases mild alcohol may be used.
Rinse and dry thoroughly.
To avoid electrical shock, do not open the housing. Refer
servicing to qualified personnel only.
Never place the equipment in water or other liquids.
RH
basic
If flammable or toxic fluids are spilled on or near the unit,
refer to your Hazardous Material Unit for cleanup and
decontamination or contact the manufacturer.
Material information for cleaning:
Hot Plate: Glass ceramic
Hot Plate frame: Aluminum die cast
Adhesive between hotplate and frame: Silicone
Housing:
Aluminum die cast
Paint on housing: Powder coated (epoxy)
Paint on Hot Plate frame: Powder coated (epoxy)
Front panel: Polyester
Control Knob: Polypropylene
Rubber foot: Neoprene
If the liquid you work with is aggressive to any of the above
materials, then clean immediately upon contact.