Manufactured to comply with the following:
EMV-regulation:
International
USA:
Canada:
France:
Germany:
Klass.
VDE 411
Technical Data
Voltage
or
Protection against
electric shock
Wire
Power consumption
Fuses (time delay) 115V:
230V:
Stirring Function
Motor
Speed range
89/336/EWG
IEC1010
UL 3101-1
CAN/CSA C 22.2
1010.1 -92
CEI 1010-1
DIN EN 61010
115 V (±10%), 60 Hz
230 V (±10%), 50/60 Hz
(see nameplate)
1 (grounded)
2 + ground
(single phase)
515 Watt
2x5 AMP(T)
2 x 2.5 AMP (T)
12VDC, brushless,
electronically commutated
150 to 1200 rpm
Heating Function
Heater
Max. surface temp
Safety cut-off
Housing
Support for H16 V
Weight:
Dimensions (LxWxH)
500 Watt
842°F (450 °C)
1202°F(650°C)
(internal heater temperature)
Thread M10
6.34 lbs. (2.88 kg)
11 x 6.7 x 3.9 (inches)
276x170x98 (mm)
Permissible Ambient Conditions
Temperature
Relative humidity
Altitude
Duration of operation
Protection class
Radio-screened
Degree of pollution
41 to104°F(5to40°C)
maximum 85 % for
temperatures up to
31
°C decreas-
ing linearly to 50% at 40°C
up to 2000 m
100%
IP 21
according to VDE 0871
2
Excess voltage category II