Operation Limitations: Do not autoclave the unit! Do not use an ultrasound bath to clean! Do not drop in water or
solvent bath. The unit is water resistant, but not water proof!
3.2 Software instructions
The NanoVolume Applications and Cuvette Applications are very similar concerning the analysis of dsDNA, ssDNA, RNA,
Oligonucleotides, protein UV and protein dye analysis. This section describes the specific features which have to be
considered using the NanoPhotometer
®
P-Class Submicroliter Cell. For general information please follow the detailed
instructions under Nanovolume Applications and Cuvette Applications.
The procedure is as follows:
Exemplary Parameter Screen
Parameter Screen
Step 1 Press 1 to select NanoVolume Applications folder
Step 2 Press 1 to select Nucleic Acids folder OR 2 to select
Protein folder.
Step 3 Select the method you want to use by pressing the
corresponding number.
Step 4 Select the Lid Factor using the left and right arrows.
Lid
Sample volume
Pathlength
Dilution
5 (optional)
3.5 5 µl
2 mm
1:5
10 (optional for P 300)
1 3 µl
1 mm
1:10
50
0.3 2 µl
0.2 mm
1:50
100 (optional)
0.3 2 µl
0.1 mm
1:100
250 (optional)
0.3 2 µl
0.04 mm
1:250
Step 5 Select subsequent parameters and specifications as
described under 4. Nanovolume Applications and
Cuvette Applications.
After the selections are confirmed the results screen displays in
top left corner the chosen Lid and the required sample volume.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen NanoPhotometer and is the answer not in the manual?

Implen NanoPhotometer Specifications

General IconGeneral
CategorySpectrophotometer
Light SourceXenon flash lamp
Measurement Time< 5 seconds
Detector2048 element CCD array
ConnectivityUSB, Ethernet, Wi-Fi
Absorbance Range0.02 - 300 A (depending on path length)