NanoPhotometer
®
P-Class User Manual
Version 2.1 Page 66 / 70
11. SOLVENT COMPATIBILITY
Most solvents typically used in life science laboratories are compatible with the NanoPhotometer® P-Class
submicroliter cell surface and the lid surfaces.
The following solvents are compatible for use on the submicroliter cell at room temperature:
- acetone (up to 5%)
- acetonitrile
- benzene
- butanol
- carbon tetrachloride
- chloroform
- ethanol
- ether
- HEPES
- hexane
- isopropanol
- MES
- methanol
- methylenchlorid
- MOPS
- phenol (up to 1%)
- n-propanol
- toluene
- buffers containing phosphate or chloride salts
e.g. PBS-buffer pH 4-10 or citrate or borate
- low concentrated acids and bases
Please note that all kind of high concentrated acids and bases are not recommended.
It is recommended to wipe off every sample immediately upon completion of a measurement.
For more specific solvents, please contact the Implen support team (support@implen.de) to check the
compatibility.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen NanoPhotometer and is the answer not in the manual?

Implen NanoPhotometer Specifications

General IconGeneral
CategorySpectrophotometer
Light SourceXenon flash lamp
Measurement Time< 5 seconds
Detector2048 element CCD array
ConnectivityUSB, Ethernet, Wi-Fi
Absorbance Range0.02 - 300 A (depending on path length)
Display7" Touchscreen

Related product manuals