NanoPhotometer
®
P-Class User Manual
Version 2.1 Page 23 / 70
Results Screen
Results Screen
Step 14 Apply/insert the reference sample. Press Blank Key. This
will be used for all subsequent samples until changed.
Step 15 Apply/insert sample and press Sample . This measures
at 260nm, 280nm, 320nm and the dye specific
wavelength and displays the result. Protein concentration
(corrected by background wavelength value if selected),
dye concentration and degree of labelling is calculated.
Step 16 If the absorbance value of the sample is not in the linear
range a “Warning message” will pop up and “Instruction”
will be displayed in the top left corner of the result screen.
Please refer 3.2 Software instructions/important
information on page 11 for further information.
Step 17 Repeat for all samples.
Step 18 Press Menu/Options to display available Options which
are described on page 8.
Step 19 Press Escape and confirm with Yes to return to the
Protein folder.
To change parameters, print or save methods press the Menu/Options button. The options menu will be opened. For
further explanation please see 2.3 Keypad and display on 6 (P 300) and 7 (P330 / P 360).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Implen NanoPhotometer and is the answer not in the manual?

Implen NanoPhotometer Specifications

General IconGeneral
CategorySpectrophotometer
Light SourceXenon flash lamp
Measurement Time< 5 seconds
Detector2048 element CCD array
ConnectivityUSB, Ethernet, Wi-Fi
Absorbance Range0.02 - 300 A (depending on path length)
Display7" Touchscreen

Related product manuals