Start-up
17
5
IMT Analytics AG
5.3.4 Flow High
4
The Flow High connector can be used for the following measurements:
• Large flows (-300 – 300nL/min)
• Volume
• Temperature
• Humidity
• Oxygen
• Pressure in the channel
Measurements can be conducted bidirectionally.
Flow High
Flow High Measuring range
Accuracy
-300 – 300 nL/min
± 1.75% of reading or ± 0.1 nL/min
Volume Measuring range
Accuracy
0 – 10 nL
± 2% of reading or ± 0.02 nL
Temperature Measuring range
Accuracy
0 – 50°C
± 1.75% of reading or ± 0.5°C
Humidity Measuring range
Accuracy
0 – 100% (non-condensing)
± 3% RH
Oxygen Measuring range
Accuracy
0 – 100%
± 1% O
2
Pressure in the
channel
Measuring range
Accuracy
0 – 150 mbar
± 0.75% of reading or ± 0.1 mbar
If work is being performed at a relatively high level of air humidity,
care must be taken to ensure that no condensation forms in the
device. Water can damage the sensors beyond repair!
4
Standard litres per minute (converted to STP conditions of 21.1°C and 1013 mbar)