Start-up
17
5
IMT Analytics AG
5.3.4 Flow High
4
The Flow High connector can be used for the following measurements:
Large flows (-300 – 300nL/min)
Volume
Temperature
Humidity
Oxygen
Pressure in the channel
Measurements can be conducted bidirectionally.
Flow High
Flow High Measuring range
Accuracy
-300 – 300 nL/min
± 1.75% of reading or ± 0.1 nL/min
Volume Measuring range
Accuracy
0 – 10 nL
± 2% of reading or ± 0.02 nL
Temperature Measuring range
Accuracy
0 – 50°C
± 1.75% of reading or ± 0.5°C
Humidity Measuring range
Accuracy
0 – 100% (non-condensing)
± 3% RH
Oxygen Measuring range
Accuracy
0 – 100%
± 1% O
2
Pressure in the
channel
Measuring range
Accuracy
0 – 150 mbar
± 0.75% of reading or ± 0.1 mbar
If work is being performed at a relatively high level of air humidity,
care must be taken to ensure that no condensation forms in the
device. Water can damage the sensors beyond repair!
4
Standard litres per minute (converted to STP conditions of 21.1°C and 1013 mbar)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IMT Analytics FlowAnalyser Series and is the answer not in the manual?

IMT Analytics FlowAnalyser Series Specifications

General IconGeneral
BrandIMT Analytics
ModelFlowAnalyser Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals