Start-up
19
5
IMT Analytics AG
5.3.7 ±1 bar pressure sensor (PF-301 VAC)
With FlowAnalyser PF-301 VAC there is an additional ± 1 bar sensor that is connect-
ed to the designated port. The connecting nipple is marked with a yellow ring.
±1 bar pressure sensor
Measuring range: -1000 – 1000 mbar
Accuracy: ± 0.5% of reading or 2 mbar
With the ± 1 bar pressure sensor variant a port of the differential
pressure sensor (± 150 mbar) is connected to the remaining con-
nector and the second one is connected to the environment. The
measuring range remains the same.