10
tina48e1-c (2017-10) MPG400-401.om
The gauge may be mounted in any orientation. To keep condensates
and particles from getting into the measuring chamber preferably
choose a horizontal to upright position and possibly use a seal with a
centering ring and filter.
If adjustment should be possible after the gauge has been installed, be
sure to install it so that the <HV> and <ATM> trimmer potentiometers
can be accessed with a screw driver ( 16).
Remove the protective lid and install the product to the vacuum system.
When making a CF flange connection, it may be advantageous to temporarily
remove the magnet unit ( 11).
or
Protective lid
Clamp
Seal with centering ring
Seal with centering
ring and filter
Keep the protective lid.
Procedure