tina48e1-c (2017-10) MPG400-401.om 25
Problem Possible cause Remedy
Measuring signal
continually < 0.5 V
"Error low".
No supply voltage. Turn on the power
supply.
Measuring signal
continually > 9.5 V
"Error high".
Pirani measurement
element defective
(filament rupture).
Replace the Pirani
element
(MPG400 20)
(MPG401 23).
Electronics unit not cor-
rectly mounted.
Mount the electronics
unit correctly
(MPG400 20)
(MPG401 24).
The green lamp is ON and
the identification indicates
Pirani-only mode (mea-
suring signal continually
> 4.0 V).
The cold cathode dis-
charge has not ignited.
Wait until the gas dis-
charge ignites (in case of
contamination with insu-
lation layers, the cold
cathode may completely
fail to ignite).
(Cleaning
MPG400 18
MPG401 21).
The MPG has only been
activated with
p < 3×10
—9
mbar.
Slightly increase the
pressure.
Measuring signal con-
tinually > 5 V or display
> 10
-3
mbar although
vacuum pressure is OK.
Pirani measurement cir-
cuit not adjusted, e.g.
due to severe contami-
nation.
Readjust the Pirani
measurement circuit
( 16). If adjustment
is impossible, replace the
Pirani element.
Measurement of heavy
gases.
Convert with the corre-
sponding formula
( 30).
Severe outgassing in the
cold cathode measuring
chamber.
Clean the measuring
chamber.
Measuring signal unstable. Gauge contaminated. Clean the gauge
(MPG400 18)
(MPG401 21).
6.4 Troubleshooting