tina48e1-c (2017-10) MPG400-401.om 7
Adjustment
<HV> potentiometer at <10
-4
mbar
<ATM> potentiometer at atmospheric pressure
Electrical connection FCC68 female, 8 poles
Sensor cable 8 conductors plus shielding
Line length
50 m (8×0.14 mm²)
Operating voltage 3.3 kV
Operating current 500 µA
Grounding concept
("Electrical Connection")
Vacuum connection-signal common
connected via 10 kΩ
(max. voltage differential
with respect to safety ±50 V
with respect to accuracy ±10 V)
Supply common-signal common conducted separately
Materials exposed to vacuum
Vacuum connection
Measuring chamber
Feedthrough
Internal seals
MPG400
MPG401
Anode
Ignition aid
Pirani measuring tube
Pirani filament
stainless steel
stainless steel
ceramic
FPM 75
Ag, Cu, soft solder (Sn, Ag)
Mo
stainless steel
Ni, Au
W
Mounting orientation any
Internal volume
20 cm³
Pressure 10 bar (absolute)
limited to inert gases
Temperatures
Operation
MPG400
MPG401
+5 … +55 °C
+5 … +150 °C
(at flange in horizontal mounting orien-
tation, without magnetic shielding)
Bakeout +150 °C
(without magnetic shielding and
electronics unit)
Pirani filament +120 °C
Storage –40 °C … +65 °C
Relative humidity 80% at temperatures up to +31 °C
decreasing to 50% at +40 °C
Use indoors only
altitude up to 2000 m
Degree of protection IP 40